Features
- Operates in 300 torr vacuum
- KF-25 flange connector
- No maintenance Zirconium cell
- No calibration required
- 3 year warranty
- No drift to environmental or
- temperature changes
- 10 + year Sensor life
- Local digital display, 4-20mA output
- Joystick adjustable dual alarm relays
- Integral computer controlled electronics
- C UL listed Measuring Equipment E363306
- Ce approved EMC EN 61326-1:2006 / EN 61010-1-3-2013 LVD
The Air check O2 Deficiency Monitor is a compact gas monitoring system that’s ideal for the
continuous monitoring of gas lines, 3D printers, nitrogen gas compressors, vacuum chambers and
locations where low oxygen levels need to be monitored. Unlike electrochemical sensor cells, the Air
check O2 zirconium cell provides stable oxygen readings even in areas where temperature and
humidity levels are changing. And, the Air check O2 is one of the only monitors suitable for
monitoring directly in a vacuum.
The heart of the monitoring system is a long lasting zirconium sensor, which detects low oxygen
conditions within seconds and provides accurate measurements over a wide temperature and humidity
range. The zirconium O2 sensor cell will operate continuously for 10 or more years and requires an
absolute minimum of maintenance. There are no zero or span calibration pots to adjust and when
compared to disposable type sensors, our long life zirconium O2 sensor can save up to $475 annually
and will pay for itself in just over 3 years!
The Air check unique sensor mounts directly into chambers or load locks through the use of a KF
type vacuum fitting. Monitoring oxygen directly eliminates the need to change the environment back to
ambient and can improve thru put. The Air check O2 Deficiency Monitor also provides stable
readings even in areas that contain high electro-magnetic fields.
Connects to DCS and PLC Controls
The Air check O2 Deficiency Monitor is 24VDC powered and transmits continuous oxygen
concentration levels to any distributive control system or programmable logic controller.
PureAire’s Oxygen Sensor Cell
The Air check O2 Deficiency Monitor uses an exclusive Current Limiting Zirconium Oxide Oxygen
sensor that never requires a reference gas. Unlike concentration type zirconium cells that must have a
reference gas, PureAire’s O2 monitor can operate in 100% nitrogen environments. Capable of
detecting 0% up to 25% oxygen levels, the current limiting O2 sensor operates at a lower temperature
than competitive concentration type cells. The average life of PureAire’s O2 sensor is over 10 years in
most environments.
The Air check O2 Deficiency Monitor never needs calibration. The earth is a wonderful source of
calibrated oxygen and under ambient levels, PureAire’s O2 monitor is continuously being challenged to
20.9%. There are no zero or span adjustments to make; the only optional response test required is to
subject the O2 system to nitrogen periodically.
O2 Monitor System Features
The Air check O2 Deficiency Monitor is available in many different configurations. PureAire uses a
sophisticated built-in CPU that is flexible to provide users with a low cost basic display only monitor or a
full featured monitor with dual level, user selectable alarm relays.